1. Measurement content:

 

1.1 Working frequency band

The working frequency band of the hydrophone is 20Hz100kHz.

 

1.2. Receiving sensitivity

The receiving sensitivity within 0.02KHz-50KHz is not less than -197dB, and other frequency bands are available.

 

1.3. Horizontal directivity

Omnidirectional, measuring frequency points 10kHz, 30kHz, 50kHz.

 

1.4. Vertical directivity

Measure the vertical directivity of the transducer, measuring frequency points 10kHz, 30kHz, 50kHz.

 

1.5. Hydrostatic pressure: 20MPa.


2.Measurement data

 

1. 

2.1 static capacitance

 


2024-01#

2024-02#

2024-03#

2024-04#

2024-05#

2024-06#

C0/nF

2.89

2.94

2.98

2.88

2.9

2.92

 

 

2.2 receiving sensitivity

f/kHz

2024-01#

2024-02#

2024-04#

2024-05#

2024-06#

0.02

-190.9

-190.6 

-190.8

-190.8

-190.4

0.025

-190.5

-190.7

-190.6 

-190.6

-190.5 

0.0315

-190.4 

-190.6

-190.6

-190.7 

-190.5

0.04

-190.4 

-191.0

-191.0

-191.0

-191.4

0.05

-190.4

-190.7 

-190.8

-190.6 

-190.4 

0.063

-190.5

-190.6 

-190.7 

-190.9

-190.6

0.08

-190.5 

-190.7 

-190.8

-190.6 

-190.4 

0.1

-189.7

-190.8

-190.6 

-191.0

-190.5

0.125

-190.4

-190.6 

-190.7 

-190.6

-191.0

0.16

-190.4 

-190.6 

-190.6

-190.6 

-190.4

0.2

-190.4 

-190.7 

-190.6 

-190.7

-190.7

0.25

-190.6

-190.6 

-190.7

-190.7

-191.0

0.315

-191.0

-190.6 

-190.8

-190.6 

-190.4 

0.4

-190.1

-190.6 

-190.7

-190.7 

-190.9

0.5

-191.0

-190.7 

-190.9

-190.6 

-190.5

0.63

-190.4 

-190.7 

-191.0

-190.8

-190.9

0.8

-190.5 

-191.1

-190.6 

-190.6

-190.9

1

-190.9

-190.7 

-190.6 

-190.6 

-190.4 

1.25

-190.6 

-190.7 

-190.7

-190.7 

-190.5

1.6

-190.6

-190.6

-190.6 

-190.6

-190.8

2

-190.8

-190.6 

-190.6 

-191.1

-191.2

2.5

-191.2

-191.2

-191.9

-191.5

-192.1

3.15

-190.8

-191.4

-191.7

-190.6 

-191.9

4

-192.2

-190.6 

-192.6

-192.5

-192.4

5

-190.9

-191.9

-191.5

-191.7

-192.3

6.3

-190.4 

-191.4

-191.2

-191.2

-191.9

8

-192.0

-190.7 

-192.2

-191.1

-192.0

10

-190.9

-191.7

-193.2

-191.4

-191.5

12.5

-191.4

-192.3 

-192.5

-192.1

-192.0

16

-191.5

-193.6 

-193.5 

-193.1

-192.5

20

-193.4 

-193.3

-193.6

-193.3

-193.4 

25

-193.4 

-193.6 

-193.9

-192.9

-193.4

31.5

-193.3

-193.5

-194.1

-193.6 

-193.4 

40

-193.3 

-193.5 

-194.8

-193.5

-193.3 

50

-193.2 

-193.6 

-193.2

-193.4 

-193.3

63

-200.3

-200.4

-200.4

-200.1

-200.3

80

-207.8

-207.9

-207.6

-207.5

-207.7

100

-215.0

-215.2

-215.3

-215.2

-215.4







1. 

1.1. 

 

2.3 Horizontal directivity

 

 


 

 

 


2.4 Vertical directivity






The 20KHz-100KHz transmitting measurements are shown in the table below:

 

 

Frequency

20KHz

30KHz

40KHz

50KHz

60KHz

70KHz

80KHz

90KHz

100KHz

voltage response

124dB

131dB

136dB

141dB

144dB

142dB

137dB

136dB

134dB

sound source level (SL)

 

184dB

191dB

192dB

194dB

195dB

194dB

194dB

194dB

191dB

 

 

Hydrostatic pressure test

After 20MPa/0.5h, 0MPa/0.25h, and 20MPa/0.5h hydrostatic pressure tests, the insulation resistance between the positive and negative electrodes of the hydrophone is greater than 100MΩ.

 

 


Feedback

1.2Khz High Power Transmitting Transducer Underwater Acoustic Transducer

Low Frequency High Power Underwater Acoustic Transducer transmitting Transducer

1.2kHz High Power Ultrasonic Transducer Underwater Acoustic Transducer 

5.7kHz Cylindrical Underwater Acoustic Transducer for Underwater Communication System

Custom 10Khz Underwater Acoustic Transducer for Marine Detection