WuHan Piezohannas Tech.Co .,Ltd is a manufactuer of piezoelectric ceramics ,ultrasonic transducers and some other ultrasonic deivce with strong technology force.With a quality management system and research & development sector,our products are extensively used in most applications.
Piezo Ceramics Description:
Dimensions: diameter 14mm,
Focal length: 16mm
Standard Size of HIFU Piezo Ceramic:
Working Frequency : 1 ~ 10 MHz (Selectable)
Aperture size : Diameter 20 ~ 70mm(Selectable)
Focal length: 10 ~ 100 mm (Selelctable)
Strong Electrode: Silvering ,Nickel, Titanium Alloys
Long Lifetime: high power pulses>15000
Typical value of "Hard" PZT materials performance:
Hard" PZT materials |
PZT-41 |
PZT-42 |
PZT-43/4D |
PZT-82 |
PBaS-4 |
||
Dielectric Constant |
ɛTr3 |
1050 |
1250 |
1420 |
1100 |
1900 |
|
Coupling factor |
KP |
0.58 |
0.58 |
0.58 |
0.52 |
0.59 |
|
K31 |
0.32 |
0.33 |
0.34 |
0.3 |
0.34 |
||
K33 |
0.66 |
0.67 |
0.68 |
0.57 |
0.68 |
||
Kt |
0.48 |
0.48 |
0.48 |
0.4 |
0.49 |
||
Piezoelectric coefficient |
d31 |
10-12m/v |
-106 |
-124 |
-138 |
-100 |
-160 |
d33 |
10-12m/v |
260 |
280 |
300 |
240 |
380 |
|
g31 |
10-3vm/n |
-11.4 |
-11.2 |
-11 |
-10.3 |
-9.5 |
|
g33 |
10-3vm/n |
28 |
25.3 |
24 |
25 |
22.6 |
|
Frequency coefficients |
Np |
2280 |
2200 |
2160 |
2280 |
2200 |
|
N1 |
1671 |
1613 |
1583 |
1671 |
1613 |
||
N3 |
1950 |
1900 |
1875 |
1950 |
1850 |
||
Nt |
2250 |
2200 |
2200 |
2300 |
2200 |
||
Elastic compliance coefficient |
Se11 |
10-12m2/n |
11.8 |
12.7 |
13.2 |
11.6 |
13.2 |
Machanical quality factor |
Qm |
1000 |
800 |
600 |
1200 |
2200 |
|
Dielectric loss factor |
Tg δ |
% |
0.3 |
0.4 |
0.5 |
0.3 |
0.5 |
Density |
ρ |
g/cm3 |
7.5 |
7.5 |
7.5 |
7.6 |
7.5 |
Curie Temperature |
Tc |
°C |
320 |
320 |
320 |
310 |
310 |
Young's modulus |
YE11 |
<109N/m3 |
85 |
79 |
76 |
86 |
76 |
Poison Ratio |
0.3 |
0.3 |
0.3 |
0.3 |
0.33 |
High Intensity Focused Ultrasound:
Ultrasound focusing has good tissue penetration and energy storage. The host generates high-frequency electrical signals, which are converted into focused ultrasound by transducer. The ultrasound is transmitted to the lesion tissues by circulating water system, sound transmission membrane and contact coupling agent. Ultrasound can produce thermal effect, mechanical effect, cavitation effect and biochemical effect in biological tissues. Polaris uses these characteristics to act on pathological tissues to degenerate them, promote tissue reconstruction and improve microcirculation to achieve therapeutic purposes.
Application Image: